Operating Pressure:
1) Offers a versatile operating pressure range from 4 to 20 cmH2O (or 4 to 20 hPa) to suit various patient needs.
2) High precision with a resolution ratio of 0.1 cmH2O (0.1 hPa) for accurate pressure delivery.
3) Supports dynamic pressure variations within a range of 0.2 cmH2O (0.2 hPa) for adaptable therapy.
Ramp Time:
1) Variable operating range from 0 to 45 minutes, allowing gradual pressure increase to enhance patient comfort.
2) Precise control with a 1-minute resolution ratio, ensuring tailored therapy sessions.
3) Dynamic variation options of 5 minutes, providing flexibility in therapy initiation.
Power:
1) Compatible input voltage range from 100 to 240 V~, 50-60 Hz; 1.5 A MAX, ensuring global usability.
2) Stable output of DC 24 V 3.75 A, delivering consistent performance.
Humidifier:
1) Adjustable volume capacity from 200ml to 360ml, catering to personalized humidification needs.
2) Efficient power consumption not exceeding 48W, promoting energy conservation.
3) Constructed with robust PC materials and a stainless steel chassis for durability and safety.
4) Heating plate reaches a maximum temperature of 63ºC (±3ºC), assuring optimal humidification.
5) Breathing tube air outlet temperature remains below 41ºC for safe and comfortable use.
6) Equipped with a thermal circuit breaker: disconnection at >85±5ºC, recovery at <55±15ºC, with a delay of <10s, DC24V-3A, for enhanced safety.
7) Protective disconnection temperature set at 76ºC to prevent overheating.
Safety Classification:
1)(Anti-electric shock level): Classified as Class II and Type BF applied part for maximum patient safety.
2) SpO2 Device: Recognized as a Class I laser product, ensuring precise and safe monitoring.
3) Ingress protection rating of IP22, providing adequate protection against environmental factors.
SpO2 Measurement:
1) Comprehensive measurement range from 0% to 100%, accurately assessing oxygen saturation levels.
2) Measurement precision with a resolution ratio of 1%, ensuring reliable readings.
3) Accuracy maintained at ±2%, delivering trustworthy SpO2 measurements.
PR Measurement:
1) Pulse Rate measurement range from 30 to 250 bpm, accommodating a wide spectrum of patient needs.
2) Precision with a 1bpm resolution ratio, ensuring accurate pulse rate tracking.
3) Accuracy of ±2 bpm or ±2% of the maximum, providing reliable pulse readings.
Noise :
1) Experience unparalleled tranquility with our device's noise output: when pressure is 10 cmH2O and distance is 1m during normal operation (including humidifier), the noise level is impressively low, at ≤35dB.
Operating Measurement:
1) Flow Sensor - Ensures precision in every breath. ±500Pa - For accurate data collection and analysis.
2) Pressure Sensor - Vital for maintaining optimal performance. 0~45 cmH2O (0~45 hPa) - Designed for a broad range of applications.